How to interpret Gage R&R results
A Gage R&R study splits the variation you see into the part of it that comes from the parts themselves and the part that comes from measuring them. The report has many numbers, but the decision comes down to a few of them.
The numbers that matter
| Metric | Compares measurement variation with… | Use it when |
|---|---|---|
| %Study Var (%SV) | the total variation in the study (as standard deviations) | the gage is used for process control / SPC |
| %Tolerance | the specification width | the gage is used to accept or reject parts |
| %Contribution | the total variance (squared units) | you want components that add up to 100% |
| ndc | how many distinct groups of parts the gage can separate | always — a resolution sanity check |
Acceptance criteria
| Verdict | %Study Var or %Tolerance (AIAG) | %Contribution (common guideline) |
|---|---|---|
| Acceptable | under 10% | under 1% |
| Conditionally acceptable — depends on the importance of the characteristic, cost of the gage, cost of repair | 10% to 30% | 1% to 9% |
| Unacceptable — improve the measurement system | over 30% | over 9% |
In addition, the AIAG MSA manual asks for ndc of 5 or more. ndc = 1.41 × (part sigma / gage sigma), rounded down; an ndc of 1 means the gage cannot tell parts apart at all.
%Contribution uses variances, %Study Var uses standard deviations, so the thresholds differ: 10% of the standard deviation is 1% of the variance, and 30% is 9%.
Worked example
The classic AIAG/Minitab study: 10 parts, 3 operators, 3 trials each, tolerance 8. The operator × part interaction is not significant (p = 0.97) and is pooled into repeatability.
| Source | %Contribution | %Study Var | %Tolerance |
|---|---|---|---|
| Total Gage R&R | 7.76 | 27.86 | 22.68 |
| Repeatability | 3.39 | 18.42 | 14.99 |
| Reproducibility | 4.37 | 20.90 | 17.01 |
| Part-to-part | 92.24 | 96.04 | 78.17 |
ndc = 4. Verdict: conditionally acceptable on both %Study Var (27.9%) and %Tolerance (22.7%), and ndc is just below 5. Reproducibility (operators) is slightly larger than repeatability (the gage itself), so the first improvement to try is the measurement method — fixturing, where on the part to measure, operator training — rather than a new gage.
Repeatability or reproducibility?
- Repeatability dominates (same operator, same part, different results): look at the gage itself — resolution, wear, clamping, the part's own form error, temperature.
- Reproducibility dominates (operators disagree): look at the method — different measuring positions, technique, reading of the scale, missing work instruction.
- Significant operator × part interaction: some operators struggle with particular parts — often a part feature that is hard to locate or a fixture that only fits one way.
Common mistakes
- Picking parts that are too similar. If all 10 parts come from one batch, part variation is small and %Study Var looks bad even for a good gage. Pick parts that span the expected process range, or judge on %Tolerance.
- Measuring in a fixed order. Randomize the order and hide the part numbers so operators don't remember earlier readings.
- Using too few parts. 10 parts × 3 operators × 2–3 trials is the standard design; with fewer parts the variance estimates are very uncertain.
- Mixing up 6σ and 5.15σ. Older studies use 5.15 standard deviations for the study variation; the AIAG 4th edition and Minitab use 6. The percentages of Study Var are the same either way; %Tolerance changes.
Run your own study for free
Paste Part / Operator / Measurement columns into the free calculator: it runs the ANOVA, pools the interaction when appropriate and reports %Study Var, %Tolerance, ndc and the AIAG verdict — matching Minitab on the example above.
Open the free Gage R&R calculator →