How to interpret Gage R&R results

A Gage R&R study splits the variation you see into the part of it that comes from the parts themselves and the part that comes from measuring them. The report has many numbers, but the decision comes down to a few of them.

The numbers that matter

MetricCompares measurement variation with…Use it when
%Study Var (%SV)the total variation in the study (as standard deviations)the gage is used for process control / SPC
%Tolerancethe specification widththe gage is used to accept or reject parts
%Contributionthe total variance (squared units)you want components that add up to 100%
ndchow many distinct groups of parts the gage can separatealways — a resolution sanity check

Acceptance criteria

Verdict%Study Var or %Tolerance (AIAG)%Contribution (common guideline)
Acceptableunder 10%under 1%
Conditionally acceptable — depends on the importance of the characteristic, cost of the gage, cost of repair10% to 30%1% to 9%
Unacceptable — improve the measurement systemover 30%over 9%

In addition, the AIAG MSA manual asks for ndc of 5 or more. ndc = 1.41 × (part sigma / gage sigma), rounded down; an ndc of 1 means the gage cannot tell parts apart at all.

%Contribution uses variances, %Study Var uses standard deviations, so the thresholds differ: 10% of the standard deviation is 1% of the variance, and 30% is 9%.

Worked example

The classic AIAG/Minitab study: 10 parts, 3 operators, 3 trials each, tolerance 8. The operator × part interaction is not significant (p = 0.97) and is pooled into repeatability.

Source%Contribution%Study Var%Tolerance
Total Gage R&R7.7627.8622.68
Repeatability3.3918.4214.99
Reproducibility4.3720.9017.01
Part-to-part92.2496.0478.17

ndc = 4. Verdict: conditionally acceptable on both %Study Var (27.9%) and %Tolerance (22.7%), and ndc is just below 5. Reproducibility (operators) is slightly larger than repeatability (the gage itself), so the first improvement to try is the measurement method — fixturing, where on the part to measure, operator training — rather than a new gage.

Repeatability or reproducibility?

Common mistakes

Run your own study for free

Paste Part / Operator / Measurement columns into the free calculator: it runs the ANOVA, pools the interaction when appropriate and reports %Study Var, %Tolerance, ndc and the AIAG verdict — matching Minitab on the example above.

Open the free Gage R&R calculator →